Summary of Study ST002892
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001773. The data can be accessed directly via it's Project DOI: 10.21228/M8J420 This work is supported by NIH grant, U2C- DK119886. See: https://www.metabolomicsworkbench.org/about/howtocite.php
| Study ID | ST002892 |
| Study Title | Folate levels in K562 cells following folate depletion |
| Study Summary | Culture of K562 cells in RPMI media containing 100 nM folic acid for up to 8 days followed my metabolomics targeting folate metabolites. All samples were harvested at the same time and the study was set up as a reverse time-course. Day 0 samples were cultured in 2,000 nM folate for the full 8 days. |
| Institute | Boston Children's Hospital, Harvard Medical School |
| Department | pathology |
| Laboratory | Kanarek Lab |
| Last Name | Kanarek |
| First Name | Naama |
| Address | Enders 1116.2, 300 Longwood Ave, Boston, MA 02115 |
| naama.kanarek@childrens.harvard.edu | |
| Phone | 6173557433 |
| Submit Date | 2023-09-28 |
| Num Groups | 6 |
| Raw Data Available | Yes |
| Raw Data File Type(s) | raw(Thermo) |
| Analysis Type Detail | LC-MS |
| Release Date | 2024-02-13 |
| Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
| Analysis ID | AN004751 |
|---|---|
| Chromatography ID | CH003583 |
| MS ID | MS004497 |
| Analysis type | MS |
| Chromatography type | Reversed phase |
| Chromatography system | Thermo Vanquish |
| Column | Phenomenex Kinetex C18 (150 x 3mm,2.6um) |
| MS Type | ESI |
| MS instrument type | Orbitrap |
| MS instrument name | Thermo Q Exactive Orbitrap |
| Ion Mode | UNSPECIFIED |
| Units | Arbitrary Units |
Chromatography:
| Chromatography ID: | CH003583 |
| Chromatography Summary: | 5 μl was injected onto a 2.6 μm, 150 x 3 mm C18 column (Phenomenex, 00F-4462-Y0) equipped with a 3.0 mm safe-guard column (Phenomenex, AJ0-8775). The column oven and autosampler tray were held at 30 °C and 4 °C, respectively. The following conditions were used to achieve chromatographic separation: buffer A was 0.1% formic acid; buffer B was acetonitrile with 0.1% formic acid. The chromatographic gradient was run at a flow rate of 0.250 ml min−1 as follows: 0–5 min: gradient was held at 5% B; 5–10 min: linear gradient of 5% to 36% B; 10.1–14.0 min: linear gradient from 36–95% B; 14.1–18.0 min: gradient was returned to 5% B. |
| Instrument Name: | Thermo Vanquish |
| Column Name: | Phenomenex Kinetex C18 (150 x 3mm,2.6um) |
| Column Temperature: | 30 |
| Flow Gradient: | 0.250 ml min−1 as follows: 0–5 min: gradient was held at 5% B; 5–10 min: linear gradient of 5% to 36% B; 10.1–14.0 min: linear gradient from 36–95% B; 14.1–18.0 min: gradient was returned to 5% B. |
| Flow Rate: | 250 uL/min |
| Solvent A: | 100% water; 0.1% formic acid |
| Solvent B: | 100% acetonitrile; 0.1% formic acid |
| Chromatography Type: | Reversed phase |