This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR001749. The data can be accessed directly via it's Project DOI: 10.21228/M8N14N This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST002802 |
Study Title | Nontargeted Serum Metabolomic Profiling of FXR-null and Wild-type Mice |
Study Summary | This study analyzed the serum metabolome of 10 FXR-knockout and 10 wild-type mice by RPLC-HRMS. |
Institute | University of North Carolina at Chapel Hill |
Department | Department of Environmental Sciences and Engineering |
Laboratory | Kun Lu |
Last Name | Hsiao |
First Name | Yun-Chung |
Address | 135 Dauer Dr. MHRC1104 |
ychsiao@live.unc.edu | |
Phone | 9842159592 |
Submit Date | 2023-07-31 |
Num Groups | 2 |
Total Subjects | 20 |
Num Males | 20 |
Raw Data Available | Yes |
Raw Data File Type(s) | raw(Thermo) |
Analysis Type Detail | LC-MS |
Release Date | 2023-08-15 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN004558 |
---|---|
Analysis type | MS |
Chromatography type | Reversed phase |
Chromatography system | Thermo Vanquish |
Column | Waters ACQUITY UPLC HSS T3 (100 x 2.1mm,1.8um) |
MS Type | ESI |
MS instrument type | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap |
Ion Mode | POSITIVE |
Units | m/z |
Chromatography:
Chromatography ID: | CH003425 |
Chromatography Summary: | The serum analytes were injected (10 μL) into a Waters Acquity UPLC HSS T3 (reverse phase C18, 100 Å, 1.8 μm, 2.1 mm × 100 mm) analytical column controlled at 40 °C, with the mobile phase composed of water (A) and acetonitrile (B) both added with 0.1% formic acid at a flow rate of 0.4 mL/min. The 15-min-gradient for chromatographic separation was set as the following: 2% B from 0-1 min; 2%-15% B from 1-3 min; 15%-50% B from 3-6 min; 50%-98% B from 6-7.5 min; 98% B held from 7.5-11.5 min; 98%-2% B from 11.5-11.6 min; and 2% B held from 11.6-15 min for a final re-equilibration. |
Instrument Name: | Thermo Vanquish |
Column Name: | Waters ACQUITY UPLC HSS T3 (100 x 2.1mm,1.8um) |
Column Temperature: | 40 |
Flow Gradient: | 2% B from 0-1 min; 2%-15% B from 1-3 min; 15%-50% B from 3-6 min; 50%-98% B from 6-7.5 min; 98% B held from 7.5-11.5 min; 98%-2% B from 11.5-11.6 min; and 2% B held from 11.6-15 min for a final re-equilibration. |
Flow Rate: | 0.4 mL/min |
Solvent A: | 100% water; 0.1% formic acid |
Solvent B: | 100% acetonitrile; 0.1% formic acid |
Chromatography Type: | Reversed phase |