Summary of Study ST000618
This data is available at the NIH Common Fund's National Metabolomics Data Repository (NMDR) website, the Metabolomics Workbench, https://www.metabolomicsworkbench.org, where it has been assigned Project ID PR000428. The data can be accessed directly via it's Project DOI: 10.21228/M8QW3H This work is supported by NIH grant, U2C- DK119886.
See: https://www.metabolomicsworkbench.org/about/howtocite.php
This study contains a large results data set and is not available in the mwTab file. It is only available for download via FTP as data file(s) here.
Study ID | ST000618 |
Study Title | Metabolomic Profiling of Follicular Fluid from Patients with Polycystic Ovary Syndrome and Hyper Response to In Vitro Fertilization Treatment |
Study Type | MS quallitative analysis |
Study Summary | The present study consisted in a metabolomic approach in follicular fluid samples from patients with polycystic ovary syndrome and hyper response to in vitro fertilization treatment. |
Institute | Sao Paulo Federal University |
Department | Department of Surgery, Division of Urology, Human Reproduction Section |
Laboratory | Center of research in Urology |
Last Name | Cordeiro |
First Name | Fernanda |
Address | Rua Embau, 231, Vila Clementino |
fernandabertuccez85@gmail.com | |
Phone | 11996667402 |
Submit Date | 2017-05-30 |
Num Groups | 3 |
Total Subjects | 81 |
Num Females | 81 |
Analysis Type Detail | LC-MS |
Release Date | 2017-11-20 |
Release Version | 1 |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN000948 |
---|---|
Analysis type | MS |
Chromatography type | Reversed phase |
Chromatography system | Ultima Quadrupole-Time of Flight - Waters Corporation |
Column | Waters Acquity HSS T3 (100 x 2.1mm,1.8um) |
MS Type | ESI |
MS instrument type | QTOF |
MS instrument name | Waters Ultima QTOF |
Ion Mode | POSITIVE |
Units | mass intensity |
Chromatography:
Chromatography ID: | CH000675 |
Chromatography Summary: | Liquid Chromatography was performed by using a chromatographer Acquity with a reverse phase column (Acquity UPCL HSS 1,8 µm, 2.1x100 mm - Waters Corporation, MA, USA), at 35 °C. The solvents for the elution were: A- H2O with formic acid 0.1% and B- ACN with formic acid 0.1%. The initial gradient was composed of 80% of A and 20% of B, during 1 minute. Next, the gradient was linearly changed for 50% of A and 50% of B, during the next minute. In the sequence, the gradient was changed for 35% of A and 65% of B during 2 minutes. The gradient was then inverted, considering 20% of A and 80% of B, during 3 minutes. For the next 10 minutes, the gradient was 3% of A and 97% of B. The initial condition was reestablished for the column washing during 1 minute. The total run time was 17 minutes and the solvents flux was 0.5 mL/min and the total amount of sample used per run was 5 µL. |
Instrument Name: | Ultima Quadrupole-Time of Flight - Waters Corporation |
Column Name: | Waters Acquity HSS T3 (100 x 2.1mm,1.8um) |
Column Temperature: | 35 °C |
Flow Rate: | 0.5 mL/min |
Solvent A: | 100% water; 0.1% formic acid |
Solvent B: | 100% acetonitrile; 0.1% formic acid |
Chromatography Type: | Reversed phase |