Combined analysis:
Analysis ID | AN003618 | AN003619 | AN003620 | AN003621 |
---|---|---|---|---|
Analysis type | MS | MS | MS | MS |
Chromatography type | Reversed phase | Reversed phase | Reversed phase | Reversed phase |
Chromatography system | Thermo Vanquish | Thermo Vanquish | Thermo Vanquish | Thermo Vanquish |
Column | Phenomenex Kinetex C18 (150 x 2.1mm,1.7um) | Phenomenex Kinetex C18 (150 x 2.1mm,1.7um) | Phenomenex kinetex C18 (150 x 2.1mm,1.7um) | Phenomenex kinetex C18 (150 x 2.1mm,1.7um) |
MS Type | ESI | ESI | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap | Orbitrap | Orbitrap |
MS instrument name | Thermo Q Exactive Orbitrap | Thermo Q Exactive Orbitrap | Thermo Q Exactive Orbitrap | Thermo Q Exactive Orbitrap |
Ion Mode | NEGATIVE | POSITIVE | NEGATIVE | POSITIVE |
Units | peak area | peak area | peak area | peak area |
MS:
MS ID: | MS003369 |
Analysis ID: | AN003618 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Resolution 70,000, scan range 65-900 m/z, maximum injection time 200 ms, microscans 2, automatic gain control (AGC) 3 x 10^6 ions, source voltage 4.0 kV, capillary temperature 320 C, and sheath gas 45, auxiliary gas 15, and sweep gas 0 (all nitrogen). Data converted to mzXML using RawConverter. Metabolites were annotated and integrated using Maven in conjunction with the KEGG database. |
Ion Mode: | NEGATIVE |
MS ID: | MS003370 |
Analysis ID: | AN003619 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Resolution 70,000, scan range 65-900 m/z, maximum injection time 200 ms, microscans 2, automatic gain control (AGC) 3 x 10^6 ions, source voltage 4.0 kV, capillary temperature 320 C, and sheath gas 45, auxiliary gas 15, and sweep gas 0 (all nitrogen). Data converted to mzXML using RawConverter. Metabolites were annotated and integrated using Maven in conjunction with the KEGG database. |
Ion Mode: | POSITIVE |
MS ID: | MS003371 |
Analysis ID: | AN003620 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Resolution 70,000, scan range 65-900 m/z, maximum injection time 200 ms, microscans 2, automatic gain control (AGC) 3 x 10^6 ions, source voltage 4.0 kV, capillary temperature 320 C, and sheath gas 45, auxiliary gas 15, and sweep gas 0 (all nitrogen). Data converted to mzXML using RawConverter. Metabolites were annotated and integrated using Maven in conjunction with the KEGG database. |
Ion Mode: | NEGATIVE |
MS ID: | MS003372 |
Analysis ID: | AN003621 |
Instrument Name: | Thermo Q Exactive Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | Resolution 70,000, scan range 65-900 m/z, maximum injection time 200 ms, microscans 2, automatic gain control (AGC) 3 x 10^6 ions, source voltage 4.0 kV, capillary temperature 320 C, and sheath gas 45, auxiliary gas 15, and sweep gas 0 (all nitrogen). Data converted to mzXML using RawConverter. Metabolites were annotated and integrated using Maven in conjunction with the KEGG database. |
Ion Mode: | POSITIVE |