Return to study ST002044 main page
MB Sample ID: SA192152
Local Sample ID: | D-49 |
Subject ID: | SU002126 |
Subject Type: | Human |
Subject Species: | Homo sapiens |
Taxonomy ID: | 9606 |
Species Group: | Mammals |
Select appropriate tab below to view additional metadata details:
Combined analysis:
Analysis ID | AN003325 | AN003326 | AN003327 | AN003328 |
---|---|---|---|---|
Analysis type | MS | MS | MS | MS |
Chromatography type | HILIC | HILIC | Reversed phase | Reversed phase |
Chromatography system | Thermo Dionex Ultimate 3000 | Thermo Dionex Ultimate 3000 | Thermo Dionex Ultimate 3000 | Thermo Dionex Ultimate 3000 |
Column | Waters Xbridge BEH HILIC (75 x 2.1mm,2.5um) | Waters Xbridge BEH HILIC (75 x 2.1mm,2.5um) | Waters Acquity BEH HSS T3 (100 x 2.1mm,1.8um) | Waters Acquity BEH HSS T3 (100 x 2.1mm,1.8um) |
MS Type | ESI | ESI | ESI | ESI |
MS instrument type | Orbitrap | Orbitrap | Orbitrap | Orbitrap |
MS instrument name | Thermo Fusion Orbitrap | Thermo Fusion Orbitrap | Thermo Fusion Orbitrap | Thermo Fusion Orbitrap |
Ion Mode | POSITIVE | NEGATIVE | POSITIVE | NEGATIVE |
Units | peak height | peak height | peak height | peak height |
MS:
MS ID: | MS003095 |
Analysis ID: | AN003325 |
Instrument Name: | Thermo Fusion Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | The electrospray ionization (ESI) source was operated for both positive mode (+) and negative (-) ionization at 120,000 resolution in MS1 mode and 30000 resolution in data-dependent MS2 scan mode. The spray voltage used for these positive and negative modes is 4000 and 35000 volts respectively. Sheath gas and auxiliary gas were set to 42 and 11 respectively. The mass scan range was 60-900 m/z, ACG (Automatic control gain) target at 100000 ions and maximum injection time was 50 ms for MS and ACG target was 20000 ions and maximum injection time 60 ms for MSMS was used. Data were processed using the IDSL.IPA software (https://cran.r-project.org/web/packages/IDSL.IPA/index.html) |
Ion Mode: | POSITIVE |
MS ID: | MS003096 |
Analysis ID: | AN003326 |
Instrument Name: | Thermo Fusion Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | The electrospray ionization (ESI) source was operated for both positive mode (+) and negative (-) ionization at 120,000 resolution in MS1 mode and 30000 resolution in data-dependent MS2 scan mode. The spray voltage used for these positive and negative modes is 4000 and 35000 volts respectively. Sheath gas and auxiliary gas were set to 42 and 11 respectively. The mass scan range was 60-900 m/z, ACG (Automatic control gain) target at 100000 ions and maximum injection time was 50 ms for MS and ACG target was 20000 ions and maximum injection time 60 ms for MSMS was used. Data were processed using the IDSL.IPA software (https://cran.r-project.org/web/packages/IDSL.IPA/index.html) |
Ion Mode: | NEGATIVE |
MS ID: | MS003097 |
Analysis ID: | AN003327 |
Instrument Name: | Thermo Fusion Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | The electrospray ionization (ESI) source was operated for both positive mode (+) and negative (-) ionization at 120,000 resolution in MS1 mode and 30000 resolution in data-dependent MS2 scan mode. The spray voltage used for these positive and negative modes is 4000 and 35000 volts respectively. Sheath gas and auxiliary gas were set to 42 and 11 respectively. The mass scan range was 60-900 m/z, ACG (Automatic control gain) target at 100000 ions and maximum injection time was 50 ms for MS and ACG target was 20000 ions and maximum injection time 60 ms for MSMS was used. Data were processed using the IDSL.IPA software (https://cran.r-project.org/web/packages/IDSL.IPA/index.html) |
Ion Mode: | POSITIVE |
MS ID: | MS003098 |
Analysis ID: | AN003328 |
Instrument Name: | Thermo Fusion Orbitrap |
Instrument Type: | Orbitrap |
MS Type: | ESI |
MS Comments: | The electrospray ionization (ESI) source was operated for both positive mode (+) and negative (-) ionization at 120,000 resolution in MS1 mode and 30000 resolution in data-dependent MS2 scan mode. The spray voltage used for these positive and negative modes is 4000 and 35000 volts respectively. Sheath gas and auxiliary gas were set to 42 and 11 respectively. The mass scan range was 60-900 m/z, ACG (Automatic control gain) target at 100000 ions and maximum injection time was 50 ms for MS and ACG target was 20000 ions and maximum injection time 60 ms for MSMS was used. Data were processed using the IDSL.IPA software (https://cran.r-project.org/web/packages/IDSL.IPA/index.html) |
Ion Mode: | NEGATIVE |