Return to study ST001892 main page
MB Sample ID: SA175951
| Local Sample ID: | 3 |
| Subject ID: | SU001970 |
| Subject Type: | Mammal |
| Subject Species: | Mus musculus |
| Taxonomy ID: | 10090 |
| Gender: | Male |
| Species Group: | Mammals |
Select appropriate tab below to view additional metadata details:
Combined analysis:
| Analysis ID | AN003072 | AN003073 |
|---|---|---|
| Chromatography ID | CH002273 | CH002273 |
| MS ID | MS002859 | MS002860 |
| Analysis type | MS | MS |
| Chromatography type | Reversed phase | Reversed phase |
| Chromatography system | Waters Acquity | Waters Acquity |
| Column | Waters Acquity BEH C18 (50 x 2.1mm,1.7um) | Waters Acquity BEH C18 (50 x 2.1mm,1.7um) |
| MS Type | ESI | ESI |
| MS instrument type | QTOF | QTOF |
| MS instrument name | Waters Synapt G2 | Waters Synapt G2 |
| Ion Mode | POSITIVE | NEGATIVE |
| Units | peak area | peak area |
MS:
| MS ID: | MS002859 |
| Analysis ID: | AN003072 |
| Instrument Name: | Waters Synapt G2 |
| Instrument Type: | QTOF |
| MS Type: | ESI |
| MS Comments: | Negative and positive electrospray ionization (ESI) data-independent modes were used for data acquisition with leucine enkephalin ([M+H]+ = 556.2771, [M-H]- = 554.2615) as Lock-Spray®. Operating conditions for ESI were: capillary voltage 2.75 kV, cone voltage 30 V, desolvation temperature 500°C, desolvation gas flow 1000 L/Hr. |
| Ion Mode: | POSITIVE |
| MS ID: | MS002860 |
| Analysis ID: | AN003073 |
| Instrument Name: | Waters Synapt G2 |
| Instrument Type: | QTOF |
| MS Type: | ESI |
| MS Comments: | Negative and positive electrospray ionization (ESI) data-independent modes were used for data acquisition with leucine enkephalin ([M+H]+ = 556.2771, [M-H]- = 554.2615) as Lock-Spray®. Operating conditions for ESI were: capillary voltage 2.75 kV, cone voltage 30 V, desolvation temperature 500°C, desolvation gas flow 1000 L/Hr. |
| Ion Mode: | NEGATIVE |