Return to study ST000317 main page
MB Sample ID: SA014478
| Local Sample ID: | SA019 |
| Subject ID: | SU000337 |
| Subject Type: | Bacterial cells |
| Subject Species: | Bacteria |
| Species Group: | Bacteria |
Select appropriate tab below to view additional metadata details:
Combined analysis:
| Analysis ID | AN000503 | AN000504 |
|---|---|---|
| Chromatography ID | CH000355 | CH000355 |
| MS ID | MS000439 | MS000440 |
| Analysis type | MS | MS |
| Chromatography type | Reversed phase | Reversed phase |
| Chromatography system | Unspecified | Unspecified |
| Column | Waters Acquity CSH C18 (100 x 2.1mm,1.7um) | Waters Acquity CSH C18 (100 x 2.1mm,1.7um) |
| MS Type | ESI | ESI |
| MS instrument type | QTOF | QTOF |
| MS instrument name | Agilent 6530 QTOF | Agilent 6550 QTOF |
| Ion Mode | POSITIVE | NEGATIVE |
| Units | Peak height | Peak height |
MS:
| MS ID: | MS000439 |
| Analysis ID: | AN000503 |
| Instrument Name: | Agilent 6530 QTOF |
| Instrument Type: | QTOF |
| MS Type: | ESI |
| Ion Mode: | POSITIVE |
| Capillary Voltage: | 3500 |
| Collision Gas: | Nitrogen |
| Dry Gas Flow: | 8 l/min |
| Dry Gas Temp: | 325 |
| Fragment Voltage: | 120 |
| Fragmentation Method: | Auto MS/MS |
| Ion Source Temperature: | 325 |
| Ion Spray Voltage: | 1000 |
| Ionization: | Pos |
| Precursor Type: | Intact Molecule |
| Reagent Gas: | Nitrogen |
| Source Temperature: | 325 C |
| Desolvation Gas Flow: | 11 l/min |
| Desolvation Temperature: | 350 C |
| Nebulizer: | 35 psig |
| Octpole Voltage: | 750 |
| Resolution Setting: | Extended Dynamic Range |
| Scan Range Moverz: | 60-1700 Da |
| Scanning Cycle: | 2 Hz |
| Scanning Range: | 60-1700 Da |
| Skimmer Voltage: | 65 |
| MS ID: | MS000440 |
| Analysis ID: | AN000504 |
| Instrument Name: | Agilent 6550 QTOF |
| Instrument Type: | QTOF |
| MS Type: | ESI |
| Ion Mode: | NEGATIVE |
| Capillary Voltage: | 3500 |
| Collision Gas: | Nitrogen |
| Dry Gas Flow: | 13 l/min |
| Dry Gas Temp: | 200 |
| Fragment Voltage: | 175 |
| Fragmentation Method: | Auto MS/MS |
| Ion Source Temperature: | 325 |
| Ion Spray Voltage: | 1000 |
| Ionization: | Neg |
| Precursor Type: | Intact Molecule |
| Reagent Gas: | Nitrogen |
| Source Temperature: | 325 C |
| Desolvation Gas Flow: | 11 l/min |
| Desolvation Temperature: | 350 C |
| Nebulizer: | 35 psig |
| Octpole Voltage: | 750 |
| Resolution Setting: | Extended Dynamic Range |
| Scan Range Moverz: | 60-1700 Da |
| Scanning Cycle: | 2 Hz |
| Scanning Range: | 60-1700 Da |
| Skimmer Voltage: | 65 |