Return to study ST003069 main page

MB Sample ID: SA332114

Local Sample ID:MXP500L-0-5711_1036867631_10_0_1_1_00_1036866394_12_R28V3
Subject ID:SU003184
Subject Type:Human
Subject Species:Homo sapiens
Taxonomy ID:9606
Gender:Male and female

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN005026 AN005027
Analysis type MS MS
Chromatography type Reversed phase Reversed phase
Chromatography system ExionLC AD chromatography system (AB Sciex, Framingham, MA, USA) ExionLC AD chromatography system (AB Sciex, Framingham, MA, USA)
Column MxP Quant 500 Column System (Biocrates Part No: 21117) MxP Quant 500 Column System (Biocrates Part No: 21117)
MS Type ESI ESI
MS instrument type Triple quadrupole Triple quadrupole
MS instrument name ABI Sciex 6500+ QTrap ABI Sciex 6500+ QTrap
Ion Mode NEGATIVE POSITIVE
Units µM µM

Chromatography:

Chromatography ID:CH003798
Chromatography Summary:Measurements were carried out using LC-MS/MS and flow injection (FIA)-MS/MS analyses on a Sciex 6500+ series mass spectrometer coupled to an ExionLC AD chromatography system (AB Sciex, Framingham, MA, USA), utilizing the Analyst 1.7.1 software with hotfix 1 (also AB SCIEX). All required standards, quality controls and eluents were included in the kit, as well as the chromatographic column for the LC-MS/MS analysis part. LC1_pos: The UHPLC autosampler and column oven are held at 10 °C and 50 °C, respectively. The injection volume was 5 µL. Eluent A is water (0.2% formic acid) and eluent B is acetonitrile (0.2% formic acid). Gradient run time of 5.8 mins. Remain at 0% B for 0.25 min, linear gradient to 12% B at 1.5 min and 17.5% B at 2.7 min. Further linear increase to 50% B at 4 mins and 100% B at 4.5 min, where it remains until 5 min. Reduction to 0% at 5.1 min and remaining at 0% until 5.8 min. Flow rate of 0.8 mL/min until 4.5 min, then increase to 1 mL/min until 4.7 min and remaining until 5.1 min. Finally, reduction to 0.8 mL/min until 5.8 min.
Instrument Name:ExionLC AD chromatography system (AB Sciex, Framingham, MA, USA)
Column Name:MxP Quant 500 Column System (Biocrates Part No: 21117)
Column Temperature:50°C
Flow Gradient:Gradient run time of 5.8 mins. Remain at 0% B for 0.25 min, linear gradient to 12% B at 1.5 min and 17.5% B at 2.7 min. Further linear increase to 50% B at 4 mins and 100% B at 4.5 min, where it remains until 5 min. Reduction to 0% at 5.1 min and remaining at 0% until 5.8 min. Flow rate of 0.8 mL/min until 4.5 min, then increase to 1 mL/min until 4.7 min and remaining until 5.1 min. Finally, reduction to 0.8 mL/min until 5.8 min.
Flow Rate:0.8 mL/min
Solvent A:100% Water; 0.2% FA
Solvent B:100% ACN; 0.2% FA
Chromatography Type:Reversed phase