Return to study ST001709 main page

MB Sample ID: SA159043

Local Sample ID:Sample_07
Subject ID:SU001786
Subject Type:Cultured cells
Subject Species:Homo sapiens
Taxonomy ID:9606
Species Group:Mammals

Select appropriate tab below to view additional metadata details:


Combined analysis:

Analysis ID AN002783 AN002784
Analysis type MS MS
Chromatography type HILIC HILIC
Chromatography system Thermo Vanquish Thermo Vanquish
Column SeQuant ZIC-HILIC (150 x 2.1mm,5um) SeQuant ZIC-HILIC (150 x 2.1mm,5um)
MS Type ESI ESI
MS instrument type Orbitrap Orbitrap
MS instrument name Thermo Q Exactive Orbitrap Thermo Q Exactive Orbitrap
Ion Mode POSITIVE NEGATIVE
Units Peak Area Peak Area

Chromatography:

Chromatography ID:CH002059
Chromatography Summary:Samples were run on a Vanquish (Thermo Scientific) UHPLC system with mobile phase A (20 mM ammonium carbonate, pH 9.7) and mobile phase B (100% Acetonitrile) at a flow rate of 150 µL/min on a SeQuant ZIC-pHILIC Polymeric column (2.1 × 150 mm 5 μm, EMD Millipore) at 35°C. Separation was achieved with a linear gradient from 20% A to 80% A in 20 min followed by a linear gradient from 80% A to 20% A from 20 min to 20.5 min. 20% A was then held from 20.5 min to 28 min.
Instrument Name:Thermo Vanquish
Column Name:SeQuant ZIC-HILIC (150 x 2.1mm,5um)
Column Temperature:35°C
Flow Gradient:100% Acetonitrile
Flow Rate:150 µL/min
Internal Standard:10 nM Trifluoromethanesulfonate (extraction buffer)
Solvent A:100% water; 20 mM ammonium carbonate, pH 9.7
Solvent B:100% acetonitrile
Chromatography Type:HILIC
  logo